Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1606M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 21.10.2017, Control date: 18.08.2016,  Control #  2

No serum blank (average) : 0.164

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.003 -0.005 0.003 -0.001 0.000 0
Negative control serum
(N° 9300-06)
0.085 0.125 0.099 0.098 0.101 4
Weak positive serum
(N° 9300-07)
0.668 0.731 0.732 0.754 0.721 30
Positive control serum
(N° 9300-08)
2.355 2.447 2.300 2.356 2.364 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.364 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 4 % Yes
Absorbance of no serum blank < 0.350 0.164 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland