Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1605D


Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 21.10.2017, Control date: 27.10.2017,  Control #  4

No serum blank (average) : 0.159

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.003 -0.001 0.003 0.001 0.000 0
Negative control serum
(N° 965
0-06)
0.133 0.172 0.123 0.125 0.138 6
Weak positive serum
(N° 965
0-07)
0.311 0.302 0.274 0.291 0.294 13
Positive control serum
(N° 9650
-08)
2.132 2.315 2.087 2.190 2.181 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.181 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 6 % Yes
Absorbance of no serum blank < 0.350 0.159 Yes

Comments:

Good for use.