Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1603D


Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 01.10.2017, Control date: 26.01.2017,  Control #  3

No serum blank (average) : 0.168

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.001 -0.012 0.003 0.008 0.000 0
Negative control serum
(N° 965
0-06)
0.114 0.109 0.107 0.127 0.114 5
Weak positive serum
(N° 965
0-07)
0.302 0.283 0.294 0.293 0.293 12
Positive control serum
(N° 9650
-08)
2.314 2.488 2.216 2.522 2.385 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.385 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.168 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch