Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1603D


Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 01.10.2017, Control date: 02.08.2016,  Control #  2

No serum blank (average) : 0.151

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.004 0.002 -0.001 0.003 0.000 0
Negative control serum
(N° 965
0-06)
0.073 0.070 0.071 0.101 0.078 4
Weak positive serum
(N° 965
0-07)
0.230 0.256 0.260 0.270 0.254 12
Positive control serum
(N° 9650
-08)
2.071 2.276 2.118 2.316 2.195 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.195 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.151 Yes

Comments:

Good for use.