Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1602B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 13.09.2017, Control date: 29.01.2016,  Control #  1

No serum blank (average) : 0.144

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.002 -0.004 0.006 -0.005 0.000 0
Negative control serum
(N° 9600-06)
0.087 0.084 0.088 0.073 0.083 4
Weak positive serum
(N° 9600-07)
0.481 0.489 0.496 0.504 0.493 23
Positive control serum
(N° 9600-08)
2.109 2.137 2.082 2.187 2.129 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.129 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.144 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch