Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1524D


Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 21.08.2017, Control date: 15.12.2016,  Control #  3

No serum blank (average) : 0.166

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.005 0.000 0.006 0.000 0.000 0
Negative control serum
(N 965
0-06)
0.058 0.073 0.081 0.081 0.073 3
Weak positive serum
(N 965
0-07)
0.265 0.263 0.236 0.273 0.259 11
Positive control serum
(N 9650
-08)
2.186 2.339 2.268 2.404 2.299 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.299 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.166 Yes

Comments:

Good for use.

Click here to read other quality controls                   Back to Home page

 