Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1524D


Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 21.08.2017, Control date: 20.06.2016,  Control #  2

No serum blank (average) : 0.157

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.000 0.000 0.000 0.001 0.000 0
Negative control serum
(N 965
0-06)
0.071 0.056 0.038 0.063 0.057 2
Weak positive serum
(N 965
0-07)
0.281 0.285 0.263 0.294 0.281 12
Positive control serum
(N 9650
-08)
2.273 2.507 2.273 2.465 2.379 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.379 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.157 Yes

Comments:

Good for use.