Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1524D


Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 21.08.2017, Control date: 05.01.2016,  Control #  1

No serum blank (average) : 0.155

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.005 -0.002 0.001 -0.003 0.000 0
Negative control serum
(N 965
0-06)
0.056 0.064 0.053 0.062 0.059 2
Weak positive serum
(N 965
0-07)
0.326 0.342 0.301 0.292 0.315 13
Positive control serum
(N 9650
-08)
2.312 2.540 2.244 2.507 2.401 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.401 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.155 Yes

Comments:

Good for use.

Click here to read other quality controls                   Back to Home page

 

Bordier Affinity Products SA.       http://www.bordier.ch/        Fax: + 41 21 633 31 78