Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1522M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 10.07.2017, Control date: 04.11.2016,  Control #  3

No serum blank (average) : 0.176

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.007 0.001 0.003 0.004 0.000 0
Negative control serum
(N 9300-06)
0.126 0.169 0.140 0.153 0.147 7
Weak positive serum
(N 9300-07)
1.050 1.025 0.994 1.028 1.024 48
Positive control serum
(N 9300-08)
2.089 2.167 2.091 2.226 2.143 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.000 2.143 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 7 % Yes
Absorbance of no serum blank < 0.350 0.176 Yes

Comments:

Good for use.