Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1522M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 10.07.2017, Control date: 10.05.2016,  Control #  2

No serum blank (average) : 0.172

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.003 0.000 -0.005 0.003 0.000 0
Negative control serum
(N 9300-06)
0.123 0.124 0.150 0.156 0.138 6
Weak positive serum
(N 9300-07)
1.065 1.074 1.152 1.114 1.101 50
Positive control serum
(N 9300-08)
2.108 2.205 2.233 2.199 2.186 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.000 2.186 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 6 % Yes
Absorbance of no serum blank < 0.350 0.172 Yes

Comments:

Good for use.