Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1522M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 10.07.2017, Control date: 18.11.2015,  Control #  1

No serum blank (average) : 0.184

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.004 0.003 -0.001 -0.005 0.000 0
Negative control serum
(N 9300-06)
0.135 0.162 0.156 0.145 0.149 6
Weak positive serum
(N 9300-07)
1.258 1.257 1.241 1.249 1.251 51
Positive control serum
(N 9300-08)
2.432 2.646 2.266 2.407 2.438 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.000 2.438 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 6 % Yes
Absorbance of no serum blank < 0.350 0.184 Yes

Comments:

Good for use.