Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Strongyloides ratti
ELISA kit for the diagnosis of human strongyloidosis

Lot 1519S

 

Results of quality control

Kit: Bordier Affinity Products SA, Strongyloides ratti ELISA. Article 9450.
Expiry date: 06.05.2017, Control date: 08.05.2017,  Control # 4

No serum blank (average) : 0.153

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.003 0.003 0.002 -0.002 0.000 0
Negative control serum
(N 9450-06)
0.103 0.106 0.109 0.101 0.104 5
Weak positive serum
(N 9450-07)
0.764 0.720 0.749 0.728 0.740 33
Positive control serum
(N 9450-08)
2.250 2.235 2.256 2.271 2.253 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.253 Yes
Absorbance of negative serum 9450-06, as % of positive serum < 12 % 5 % Yes
Absorbance of no serum blank < 0.350 0.153 Yes

Comments:

Good for use.

Click here to read other quality controls                   Back to Home page

 

Bordier Affinity Products SA.       http://www.bordier.ch/        Fax: + 41 21 633 31 78