Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1517B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 02.04.2017, Control date: 29.01.2016,  Control #  2

No serum blank (average) : 0.147

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.005 -0.003 -0.001 -0.002 0.000 0
Negative control serum
(N 9600-06)
0.129 0.120 0.114 0.113 0.119 5
Weak positive serum
(N 9600-07)
0.463 0.507 0.475 0.469 0.478 21
Positive control serum
(N 9600-08)
2.163 2.452 2.209 2.343 2.291 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.291 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.147 Yes

Comments:

Good for use.

Click here to read other quality controls                      Back to Home page

 

Bordier Affinity Products SA.       http://www.bordier.ch/        Fax: + 41 21 633 31 78