Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1514M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 10.02.2017, Control date: 08.12.2015,  Control #  2

No serum blank (average) : 0.182

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.004 0.004 -0.009 0.008 0.000 0
Negative control serum
(N 9300-06)
0.101 0.113 0.077 0.111 0.101 5
Weak positive serum
(N 9300-07)
0.969 0.996 0.910 0.945 0.955 47
Positive control serum
(N 9300-08)
2.024 2.170 1.910 1.961 2.017 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.000 2.017 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 5 % Yes
Absorbance of no serum blank < 0.350 0.182 Yes

Comments:

Good for use.

Click here to read other quality controls                   Back to Home page

Bordier Affinity Products SA.       http://www.bordier.ch/        Fax: + 41 21 633 31 78