Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Acanthocheilonema viteae (Onchocercidae)
ELISA kit for the diagnosis of human filariasis

Lot 1510V

Results of quality control

Kit: Bordier Affinity Products SA, Acanthocheilonema viteae ELISA, Article 9400.
Expiry date: 27.12.2016, Control date: 26.10.2015,  Control #  2

No serum blank (average) : 0.153

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.002 -0.004 0.006 -0.001 0.000 0
Negative control serum
(N 9400-06)
0.075 0.087 0.094 0.084 0.085 4
Weak positive serum
(N 9400-07)
0.737 0.736 0.761 0.756 0.748 33
Positive control serum
(N 9400-08)
2.183 2.291 2.270 2.350 2.274 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.274 Yes
Absorbance of negative serum 9400-06, as % of positive serum < 12 % 4 % Yes
Absorbance of no serum blank < 0.350 0.153 Yes

Comments:

Good for use.

Click here to read other quality controls                   Back to Home page

 

Bordier Affinity Products SA.       http://www.bordier.ch/        Fax: + 41 21 633 31 78