Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1509B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 27.12.2016, Control date: 28.12.2016,  Control #  4

No serum blank (average) : 0.164

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.003 0.002 -0.003 0.004 0.000 0
Negative control serum
(N 9600-06)
0.072 0.069 0.049 0.055 0.061 3
Weak positive serum
(N 9600-07)
0.363 0.363 0.326 0.364 0.354 18
Positive control serum
(N 9600-08)
1.954 2.075 1.914 1.993 1.984 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 1.984 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.164 Yes

Comments:

Good for use.

Click here to read other quality controls                      Back to Home page

 

Bordier Affinity Products SA.       http://www.bordier.ch/        Fax: + 41 21 633 31 78