Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1509B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 27.12.2016, Control date: 22.04.2016,  Control #  3

No serum blank (average) : 0.161

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.003 0.000 0.001 -0.005 0.000 0
Negative control serum
(N 9600-06)
0.074 0.072 0.048 0.051 0.062 3
Weak positive serum
(N 9600-07)
0.388 0.374 0.371 0.366 0.375 19
Positive control serum
(N 9600-08)
2.039 1.995 1.962 2.072 2.017 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.017 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.161 Yes

Comments:

Good for use.

Click here to read other quality controls                      Back to Home page

 

Bordier Affinity Products SA.       http://www.bordier.ch/        Fax: + 41 21 633 31 78