Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1508M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 15.12.2016, Control date: 16.12.2016,  Control #  4

No serum blank (average) : 0.172

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.005 -0.011 0.020 -0.005 0.000 0
Negative control serum
(N 9300-06)
0.131 0.090 0.120 0.123 0.116 6
Weak positive serum
(N 9300-07)
0.939 0.934 0.905 0.824 0.901 43
Positive control serum
(N 9300-08)
2.003 2.119 2.017 2.153 2.073 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.000 2.073 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 6 % Yes
Absorbance of no serum blank < 0.350 0.172 Yes

Comments:

Good for use.

Click here to read other quality controls                   Back to Home page