Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1508M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 15.12.2016, Control date: 11.04.2015,  Control #  3

No serum blank (average) : 0.177

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.005 -0.008 0.011 0.003 0.000 0
Negative control serum
(N 9300-06)
0.093 0.103 0.113 0.107 0.104 5
Weak positive serum
(N 9300-07)
0.914 0.954 0.929 0.893 0.922 43
Positive control serum
(N 9300-08)
2.037 2.244 2.062 2.144 2.122 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.000 2.122 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 5 % Yes
Absorbance of no serum blank < 0.350 0.177 Yes

Comments:

Good for use.

Click here to read other quality controls                   Back to Home page